M550 Double Beam SPF Spectrophotometer
This system is for measuring the sun blocking properties of fabrics and fully conforms to
European Standard EN 13758-1, American (AATCC Test Method 183.1998), British (BS 7914:
1998) and Australian and New Zealand (AS/NZS 4399:1996) test methods. The appropriate
method should be read in full by those intending to make formal measurements.
The system comprises of a CamSpec M550 double beam scanning SPF spectrophotometer, controlled by a PC with CamSpec SPF Windows™ software and an integrating sphere installed. The system is both versatile and easy to use.
This instrument can be used as a conventional double beam scanning spectrophotometer by removing the sphere and reference attenuator.
The Sun UV Penetration system for fabrics has been designed specifically for analysts working in the fabric industry, who need an accurate and reliable method of determining the sun blocking properties of fabrics for the important UV-A and UV-B wavelengths
- Highly efficient integrating sphere
- Sample-effect compensation
- Low noise, high accuracy
- User friendly operation
- Automatic calculations
- Report generation
- Validation
- Removable integrating sphere
The purpose designed integrating sphere, which meets the requirements of international standards, efficiently collects all the scattered radiation transmitted by the fabric. The Opdima surface has a very high reflectivity throughout the UV-R region, diffusely reflecting the radiation on to a solid state detector mounted at the top of the sphere.
Technical Specifications for the M550 SPF UV/VIS Spectrophotometer | ||
Display | 1/4 VGA 320x240pixels backlit LED | |
Light sources | Tungsten-Halogen and Deuterium | |
Monochromator | Littrow type with 1200 lines/mm grating | |
Scan Speed | Up to 2500 nm/min (return 3000nm/min) | |
Detectors | 2 x Silicon Photodiodes | |
Wavelength Range | 190 - 1100nm | |
Wavelength Accuracy | ± 0.3nm | |
Wavelength Resolution | 0.1nm | |
Wavelength Repeatability | ± 0.05nm | |
Baseline Flatness | ± 0.002 (After 1 Hr warm up 1100 – 200nm) | |
Zero Drift | < 0.001A per hour after warm-up | |
Noise | ± 0.0001A @ 500nm 0A | |
Bandpass | 1.5nm | |
Stray Light (A.S.T.M.) | KCl>2A at 198nm <0.05%T @ 220nm & 340nm | |
Photometric Accuracy | ± 0.5% at 0.5A, 1.0A and 2.0A | |
Photometric Range | -0.3 to 3A, 0 to 200%T, 0 to 9999 Conc | |
Printer Interface | Parallel for A4 HP and Epson printers | |
Computer Interface | USB Port | |
Power Requirements | 110/120V, 220/230V, 50/60 Hz, 110VA | |
Dimensions | 630 x 410 x 280mm packed 770 x 560 x 410 | |
Weight | 24kg packed 28Kg |
View the PDF below to see more information on the SPF instrument. Please enquire if your requirements are not listed.
Click above to open the document to see the full specification
We are an ISO 9001:2015 registered company.
click here to see our ISO certificate